Answers are grounded in this paper and cite their sources.
A systematic analysis of the time-of-flight (TOF) method for measuring the distribution of speeds in molecular beams is presented. The conditions required for resolving a given beam speed distribution are evaluated by calculating the TOF signal for a finite open time τ of the beam shutter. The halfwidth of the TOF signal Δt exceeds the ideal half width Δt0(for τ→0) by not more than 2.5%, if the resolution R=Δt/τ is equal to or larger than 5. The peak signal amplitude Um is not less than 0.8 U0/R, where U0 is the signal corresponding to the unchopped beam (τ→∞). Design criteria for beam chopper and detection system are discussed in terms of the resolution, the range of speeds to be measured, and the optimum signal-to-noise ratio. A TOF system used for the detection of high-intensity molecular beams with large speed ratio, as well as for low-intensity scattered beams, is described and typical experimental results are presented.
No takes yet. Share an insight, caveat, or question.
Hagena et al. (1968) studied this question.