Key result
Same-procedure bailout redo-TAV appears successful after index TAVR failure.
Why the study?
Infolding, deformation, and embolization are forms of valve failure linked to significant complications.
Case Report (n=1)
Early detection of misloading and infolding during TAVR is crucial, and same-procedure bailout redo-TAV can successfully rescue structural valve failure.
Answers are grounded in this paper and cite their sources.
May support bailout redo-TAV feasibility in acute failure; hypothesis-generating and should not change practice.
OBJECTIVES: Infolding, deformation, and embolization are forms of valve failure linked to significant complications. We report a successful same-procedure bailout redo-transcatheter aortic valve placement (redo-TAV) approach after failure of the index transcatheter aortic valve replacement. KEY STEPS: An Evolut 34-mm valve showed infolding with vertical-line sign after deployment. Postdilatation with a NuMED 25-mm balloon caused deformation, embolization, and significant aortic regurgitation. Bailout redo-TAV rescue with a second Evolut 34-mm valve was positioned around 1-mm deeper. Final postdilatation with a NuMED 26-mm balloon resolved paravalvular leak and gradients and improved the transcatheter heart valve configuration. COMPLICATION PREVENTION STRATEGY: Careful verification of proper Evolut valve loading prior to insertion-along with timely recapture, resheathing, and retrieval before final deployment when abnormalities are identified-could have prevented this complication. Tall-in-tall redo-TAV may impair coronary reaccess and increase coronary obstruction risk. TAKE-HOME MESSAGE: Early detection of misloading and infolding is essential to prevent severe transcatheter heart valve deformation and may help avoid the complications related in the case.
No takes yet. Share an insight, caveat, or question.
Go et al. (2026) conducted a case report in Structural valve failure during transcatheter aortic valve replacement (n=1). Bailout redo-transcatheter aortic valve placement was evaluated. Bailout redo-TAV with a second 34-mm valve successfully resolved paravalvular leak and gradients after index valve infolding, deformation, and embolization.
Synapse has enriched 10 related papers. Consider them for comparative context: