Key result
Excessive ventricular pacing occurs in ~60% of device checks, enabling corrective reprogramming in most.
Why the study?
Does an institutional improvement program identify and reduce excessive ventricular pacing in patients with cardiac electronic devices?
Cross-Sectional (n=77)
No
Does an institutional improvement program identify and reduce excessive ventricular pacing in patients with cardiac electronic devices?
Answers are grounded in this paper and cite their sources.
A significant proportion of patients with pacemakers or ICDs have unnecessary excessive ventricular pacing (>40%), which can be identified and often corrected through regular device follow-up and institutional improvement programs.
BACKGROUND: Excessive ventricular pacing is known to be detrimental. The purpose of this study was to assess ventricular pacing in the setting of an institutional improvement program in order to decrease unnecessary pacing. METHOD: This cross-sectional single-center study performed in a university hospital assessed 80 consecutive patients attending for a cardiac electronic device (pacemaker or cardioverter defibrillator) check. Forty percent of ventricular pacing was set as the cutoff level beyond which pacing was considered excessive. RESULTS: Three patients were excluded. Forty-six (59.7%) patients (group 1) had more than 40% ventricular pacing and 31 (40.3%) patients (group 2) showed ventricular pacing less than 41%. In group 1, corrective action was successful in 27 (58.7%) patients, but 19 (41.3%) continued to have ventricular pacing over 40% and were discussed accordingly. An improvement program was established at the institution in order to decrease unnecessary ventricular pacing. CONCLUSION: Unnecessary ventricular pacing was encountered in many of the patients in this study, corrective actions were performed, and an institutional improvement project was set up as a consequence.
No takes yet. Share an insight, caveat, or question.
Kossaify et al. (2012) conducted a cross-sectional in Patients with cardiac electronic devices (pacemaker or ICD) (n=77). Device follow-up and reprogramming was evaluated on Excessive ventricular pacing (>40%). Excessive ventricular pacing was observed in 59.7% of patients during routine device follow-up, and corrective reprogramming was successfully performed in 58.7% of these cases.
Synapse has enriched 10 related papers. Consider them for comparative context: