Key result
Kardia Mobile 6L and Istel HR-2000 match standard 12-lead ECG for AF detection.
Why the study?
Mobile, portable ECG recorders allow out-of-hospital heart rhythm assessment, but their effectiveness had not been tested in everyday practice.
Do portable ECG recorders (Kardia Mobile 6L and Istel HR-2000) accurately detect atrial fibrillation and sinus rhythm compared to standard 12-lead ECG in stable cardiology patients?
Observational (n=98)
Single-blind
No
Do portable ECG recorders (Kardia Mobile 6L and Istel HR-2000) accurately detect atrial fibrillation and sinus rhythm compared to standard 12-lead ECG in stable cardiology patients?
Absolute Event Rate: 22% vs 22%
p-value: p=0.65
Portable 6-lead ECG devices provide sufficient sensitivity and specificity for detecting atrial fibrillation and sinus rhythm in clinical practice, though standard 12-lead ECG remains necessary for precise conduction and repolarization measurements.
May support portable ECGs for outpatient AF screening; leaves open need for prospective validation vs 12-lead ECG.
Aims Mobile, portable ECG-recorders allow the assessment of heart rhythm in out-of-hospital conditions and may prove useful for monitoring patients with cardiovascular diseases. However, the effectiveness of these portable devices has not been tested in everyday practice. Methods and results A group of 98 consecutive cardiology patients [62 males (63%), mean age 69 ± 12.9 years] were included in an academic care centre. For each patient, a standard 12-lead electrocardiogram (SE), as well as a Kardia Mobile 6L (KM) and Istel (IS) HR-2000 ECG were performed. Two groups of experienced physicians analysed obtained recordings. After analysing ECG tracings from SE, KM, and IS, quality was marked as good in 82%, 80%, and 72% of patients, respectively (P < 0.001). There were no significant differences between devices in terms of detecting sinus rhythm [SE (60%, n = 59), KM (58%, n = 56), and IS (61%, n = 60); SE vs. KM P = 0.53; SE vs. IS P = 0.76) and atrial fibrillation [SE (22%, n = 22), KM (22%, n = 21), and IS (18%, n = 18); (SE vs. KM P = 0.65; SE vs. IS = 0.1)]. KM had a sensitivity of 88.1% and a specificity of 89.7% for diagnosing sinus rhythm. IS showed 91.5% and 84.6% sensitivity and specificity, respectively. The sensitivity of KM in detecting atrial fibrillation was higher than IS (86.4% vs. 77.3%), but their specificity was comparable (97.4% vs. 98.7%). Conclusion Novel, portable devices are useful in showing sinus rhythm and detecting atrial fibrillation in clinical practice. However, ECG measurements concerning conduction and repolarization should be clarified with a standard 12-lead electrocardiogram.
No takes yet. Share an insight, caveat, or question.
Krzowski et al. (2021) conducted an observational in Cardiovascular disease (n=98). Kardia Mobile 6L and Istel HR-2000 vs. Standard 12-lead ECG was evaluated on Atrial fibrillation detection rate (Kardia Mobile vs Standard ECG) (p=0.65). Kardia Mobile 6L and Istel HR-2000 effectively detected atrial fibrillation with no significant difference in detection rates compared to standard 12-lead ECG (22% vs 22%, p=0.65 for Kardia Mobile).