The maximum charge injection without electrode solution decomposition for activated iridium wire electrodes in bicarbonate buffered saline was 2.1 and 1.0 mC/cm/sup 2/ geometric for anodic-first and cathodic-first, respectively, 0.2-ms balanced-charge biphasic current pulses. Electrodes biased at +0.8 V vs. SCE (saturated calomel electrode) accepted charge up to 3.5 mC/cm/sup 2/ geometric with monophasic cathodal pulses.< <ETX xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">></ETX>
No takes yet. Share an insight, caveat, or question.
Beebe et al. (1988) studied this question.