Key result
Quantitative assessment of global longitudinal strain using a standardized software algorithm showed strong correlation (r=0.86) between echocardiography and cardiac magnetic resonance imaging.
Why the study?
Does a standardized software algorithm provide consistent measurements of global longitudinal and circumferential strain across echocardiography and CMR?
Cross-Sectional (n=70)
No
Does a standardized software algorithm provide consistent measurements of global longitudinal and circumferential strain across echocardiography and CMR?
A standardized software algorithm allows reliable, modality-independent assessment of global longitudinal strain across echocardiography and CMR, supporting its use in clinical routine.
Standardized algorithms may enable consistent GLS across modalities; leaves open prospective validation before routine clinical interchangeability.
Mean GLS values were -16.2±5.3% and -17.3±5.3% for echocardiography and CMR, respectively. GLS did not differ significantly between the two imaging modalities, which showed strong correlation (r=0.86), a small bias (-1.1%) and narrow 95% limits of agreement (LOA, ±5.4%). Mean GCS values were -17.9±6.3% and -24.4 ±7.8% for echocardiography and CMR, respectively. GCS was significantly underestimated by echocardiography (p<0.001). A weaker correlation (r=0.73), a higher bias (-6.5%) and wider LOA (±10.5%) were observed for GCS. GLS showed a strong correlation (r=0.92) when image quality was good, while correlation dropped to r=0.82 with poor acoustic windows in echocardiography. GCS assessment revealed only a strong correlation (r=0.87) when echocardiographic image quality was good. No significant differences for GLS between two different echocardiographic vendors could be detected. Conclusions Quantitative assessment of GLS using a standardized software algorithm allows the direct comparison of values acquired irrespective of the imaging modality. GLS may therefore serve as a reliable parameter for the assessment of global left ventricular function in clinical routine besides standard evaluation of the ejection fraction.
No takes yet. Share an insight, caveat, or question.
Riffel et al. (2015) reported a cross-sectional. Echocardiography with standardized feature tracking algorithm vs. Cardiac magnetic resonance imaging (CMR) with standardized feature tracking algorithm was evaluated on Global longitudinal strain (GLS) (r=0.86). Quantitative assessment of global longitudinal strain using a standardized software algorithm showed strong correlation (r=0.86) between echocardiography and cardiac magnetic resonance imaging.