Key result
Hydrostatic pressure-corrected 3D-CFR correlates strongly with Doppler measurements and boosts diagnostic specificity to ~92%.
Why the study?
To develop a method for calculating coronary flow reserve derived from 3D coronary angiographic parameters and intracoronary pressure data during FFR measurement.
Does 3D-derived coronary flow reserve with hydrostatic pressure correction correlate with intracoronary Doppler measurements in patients with intermediate coronary stenosis?
Cross-Sectional (n=16)
No
Does 3D-derived coronary flow reserve with hydrostatic pressure correction correlate with intracoronary Doppler measurements in patients with intermediate coronary stenosis?
Effect estimate: r = 0.89
p-value: p=<0.0001
3D-derived coronary flow reserve with hydrostatic pressure correction provides a reliable alternative to Doppler measurements for assessing coronary microvascular function during routine FFR.
Purpose: To develop a method of coronary flow reserve (CFR) calculation derived from three-dimensional (3D) coronary angiographic parameters and intracoronary pressure data during fractional flow reserve (FFR) measurement. Methods: Altogether 19 coronary arteries of 16 native and 3 stented vessels were reconstructed in 3D. The measured distal intracoronary pressures were corrected to the hydrostatic pressure based on the height differences between the levels of the vessel orifice and the sensor position. Classical fluid dynamic equations were applied to calculate the flow during the resting state and vasodilatation based on morphological data and intracoronary pressure values. 3D-derived coronary flow reserve (CFRp-3D) was defined as the ratio between the calculated hyperemic and the resting flow and was compared to the CFR values simultaneously measured by the Doppler sensor (CFRDoppler). Results: Haemodynamic calculations using the distal coronary pressures corrected for hydrostatic pressures showed a strong correlation between the individual CFRp-3D values and the CFRDoppler measurements (r = 0.89, p < 0.0001). Hydrostatic pressure correction increased the specificity of the method from 46.1% to 92.3% for predicting an abnormal CFRDoppler < 2. Conclusions: CFRp-3D calculation with hydrostatic pressure correction during FFR measurement facilitates a comprehensive hemodynamic assessment, supporting the complex evaluation of macro-and microvascular coronary artery disease.
No takes yet. Share an insight, caveat, or question.
Tar et al. (2022) conducted a cross-sectional in Intermediate epicardial coronary stenosis (n=16). 3D-derived coronary flow reserve (CFRp-3D) with hydrostatic pressure correction vs. Intracoronary Doppler measurement (CFRDoppler) was evaluated on Correlation between CFRp-3D and CFRDoppler (r = 0.89, p=<0.0001). 3D-derived coronary flow reserve with hydrostatic pressure correction strongly correlated with invasive Doppler measurements (r = 0.89, p < 0.0001) and increased specificity for predicting abnormal CFR to 92.3%.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: