The dual-rotating-retarder configuration is one of the most common forms of the Mueller matrix polarimeter. I perform an optimization of this polarimeter configuration by minimizing the condition number of the system data reduction matrix. I find the optimum retardance for the rotating retarders to be 127 degrees. If exactly 16 intensity measurements are used for a Mueller matrix calculation, a complex relationship exists between the condition number and the sizes of the angular increments of the two retarders. If many intensity measurements are made, thus overspecifying the calculation, I find broad optimal ranges of angular increments of the two retarders that yield essentially equal performance. Experimental results are given.
No takes yet. Share an insight, caveat, or question.
Matthew H. Smith (2002) studied this question.