A New Echo-Doppler Method of Assessing Pulmonary Arterial Wave Reflection: In Vivo Validation

TYTomohiko YoshidaTUTokuhisa UejimaSKSyunta Komeda

Key Result

The noninvasive echo-Doppler method for assessing pulmonary arterial wave reflection agreed well with invasive catheter measurements (bias 0.13) in a canine model of pulmonary hypertension.

Key Points

  • This research aims to validate a new method for assessing pulmonary arterial wave reflection using Doppler echocardiography.
  • Acquired Doppler echocardiographic images of tricuspid regurgitation and right ventricular outflow in a canine model of pulmonary hypertension (n=8).
  • Simultaneously measured PA pressure and flow using a dual sensor-tipped catheter under various hemodynamic conditions.
  • Conducted wave separation analysis on echo-Doppler and catheter-derived waveforms to calculate the reflection coefficient.
  • The reflection coefficient by echo-Doppler showed good agreement with catheter-derived indices (bias = 0.13, 95% limits of agreement = -0.25 to 0.26).
  • Negative correlation found between reflection coefficient and pulmonary arterial compliance.
  • Negative correlation also observed between reflection coefficient and right ventricular systolic function.

Structured PICO

Does noninvasive Doppler echocardiography accurately estimate pulmonary arterial wave reflection compared to invasive catheter measurements in a canine model of pulmonary hypertension?

P
Population
8 healthy female Beagle dogs (aged 4-5 years old, 10-13 kg) in a canine model of pulmonary hypertension induced by pulmonary arterial microembolization.
I
Intervention
Noninvasive estimation of pulmonary arterial pressure and flow waveforms using Doppler echocardiography (tricuspid regurgitation and right ventricular outflow tract flow) to quantify wave reflection.
C
Comparator
Direct invasive measurements using a dual sensor-tipped pressure and flow wire catheter.
O
Outcome
Agreement of wave reflection coefficient (RC) between echo-Doppler and catheter-derived measurements.surrogate

A novel echo-Doppler method accurately estimates pulmonary arterial wave reflection noninvasively, providing a potential flow-independent measure of right ventricular pulsatile load for early pulmonary hypertension diagnosis.

Main Result

Effect estimate: bias 0.13 (95% CI -0.25 to 0.26)

Limitations

  • The study used an animal model of pulmonary hypertension created by embolizing peripheral pulmonary arteries, which may not reflect all clinical scenarios.
  • It remains to be investigated whether the method can detect alterations in the magnitude and timing of wave reflection caused by changes in the site of narrowing.
  • The method assumes right atrial pressure is constant during systole, which may not apply to patients with severe tricuspid regurgitation.
  • The method assumes a constant RR interval, making it inapplicable to patients with large heart rate variability such as atrial fibrillation.
  • Echocardiographic estimates of pulmonary arterial systolic pressure can be challenging in clinical practice due to difficulty in obtaining clear spectral Doppler envelopes.
  • Assumes right atrial pressure is constant during systole, limiting applicability in patients with severe tricuspid regurgitation.
  • Assumes constant RR interval, limiting applicability in patients with large heart rate variability such as atrial fibrillation.
  • Requires validation for feasibility and accuracy in human patients before clinical application.

Abstract

Abstract Background: Pulmonary arterial (PA) wave reflection provides additional information for assessing right ventricular afterload, but its applications is hampered by the need for invasive pressure and flow measurements. We tested the hypothesis that PA pressure and flow waveforms estimated by Doppler echocardiography could be used to quantify PA wave reflection. Methods: Doppler echocardiographic images of tricuspid regurgitation and right ventricular outflow tract flow used to estimate PA pressure and flow waveforms were acquired simultaneously with direct measurements with a dual sensor-tipped catheter under various hemodynamic conditions in a canine model of pulmonary hypertension (n=8). Wave separation analysis was performed on echo-Doppler derived as well as catheter derived waveforms to separate PA pressure into forward (Pf) and backward (Pb) pressures and derive wave reflection coefficient (RC) defined as Pb divided by Pf. Results: RC by echo-Doppler agreed well with RC indices by catheter (RC: bias = 0.13, 95% limits of agreement = -0.25 to 0.26). RC correlated negatively with pulmonary arterial compliance and right ventricular systolic function. Conclusions: This echo-Doppler method yields accurate measurement of reflected wave in the pulmonary circulation, paving the way to a more integrative assessment of pulmonary hemodynamics in the clinical setting.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Yoshida et al. (2021) studied Pulmonary hypertension (n=8). Echo-Doppler method vs. Dual sensor-tipped catheter was evaluated on Reflection coefficient (RC) agreement between echo-Doppler and catheter (bias 0.13, 95% CI -0.25 to 0.26). The noninvasive echo-Doppler method for assessing pulmonary arterial wave reflection agreed well with invasive catheter measurements (bias 0.13) in a canine model of pulmonary hypertension.

synapsesocial.com/papers/6a1c404b67ec8852715ef27fhttps://doi.org/10.21203/rs.3.rs-412380/v1
Ask AI
Helpful
Bookmark
Share
View Full Paper