Key result
Paced left ventricular dyssynchrony was independently associated with nonresponse to cardiac resynchronization therapy at 6 months (OR 9.74; 95% CI 2.8-33.9; P<0.0001).
Why the study?
It was unknown whether pacing-induced electrical dyssynchrony at the time of CRT device implantation is associated with chronic CRT response.
Does paced left ventricular dyssynchrony prevent left ventricular reverse remodeling in heart failure patients receiving a CRT device?
Observational (n=69)
Does paced left ventricular dyssynchrony prevent left ventricular reverse remodeling in heart failure patients receiving a CRT device?
Intraoperative measurement of paced left ventricular dyssynchrony provides significant incremental value for predicting lack of left ventricular reverse remodeling after CRT implantation.
Answers are grounded in this paper and cite their sources.
May aid CRT nonresponder identification; observational data leaves open whether optimization improves remodeling rates.
Introduction We investigated whether pacing‐induced electrical dyssynchrony at the time of cardiac resynchronization therapy (CRT) device implantation was associated with chronic CRT response. Methods and Results We included a total of 69 consecutive heart failure patients who received a CRT device. Left (LVp‐RVs) and right (RVp‐LVs) pacing‐induced interlead delays were measured intraoperatively and used to determine if there was paced left ventricular (LV) dyssynchrony, defined as present when LVp‐RVs is larger than RVp‐LVs. CRT response was defined as a reduction in LV end‐systolic volume ≥15%, 6 months after implantation. Paced left ventricular dyssynchrony (PLVD) was associated with ischemic cardiomyopathy (ICM) ( χ 2 : 8; P = .005) but not with QRS morphology nor with pacing lead positions. In a univariate analysis, PLVD (odds ratio [OR], 6.53; 95% confidence interval [CI], 2.2‐18.9; P = .001), atypical left bundle branch block (LBBB) (OR, 3.3; 95% CI, 1.2‐9.4; P = .022), and ICM (OR, 5.2; 95% CI, 1.6‐17; P = .006) were associated with nonresponse. In a multivariate analysis, both PLVD (OR, 9.74; 95% CI, 2.8‐33.9; P < .0001) and atypical LBBB (OR, 5.6; 95% CI, 1.5‐20.3; P = .009) were independently associated with nonresponse. Adding PLVD to a model based on QRS morphology provided a significant and meaningful incremental value to predict LV reverse remodeling after CRT ( χ 2 to enter: 8; P < .005). Computer simulations corroborate these findings by showing that, while intrinsic electrical dyssynchrony is a prerequisite, the level of pacing‐induced dyssynchrony modulates acute CRT response. Conclusion In addition to the intrinsic electrical substrate, PLVD is strongly associated with less LV reverse remodeling, demonstrating that measuring the electrical substrate during pacing has additional value for prediction of CRT response in an already well‐selected patient population.
No takes yet. Share an insight, caveat, or question.
Gauthey et al. (2020) conducted an observational in Heart failure (n=69). Paced left ventricular dyssynchrony (PLVD) vs. Absence of PLVD was evaluated on CRT nonresponse (failure to achieve reduction in LV end-systolic volume ≥15% at 6 months) (OR 9.74, 95% CI 2.8-33.9, p=<0.0001). Paced left ventricular dyssynchrony was independently associated with nonresponse to cardiac resynchronization therapy at 6 months (OR 9.74; 95% CI 2.8-33.9; P<0.0001).
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: