The simplified Bernoulli's method significantly underestimated pulmonary transvalvular pressure drop at peak systole compared to the general Bernoulli's equation (ratio 0.82).
Observational (n=24)
Yes
Repaired congenital heart defects and healthy volunteers (n=24)
Simplified Bernoulli's Equation (SBE) vs General Bernoulli's Equation (GBE)
Pulmonary transvalvular pressure drop at peak systole (mmHg) — Ratio 0.82 (0.700-0.943), p=0.0042
Effect estimate: Ratio 0.82 (95% CI 0.700-0.943)
Absolute Event Rate: 35.47% vs 39.51%
p-value: p=0.0042
PURPOSE: To determine whether neglecting the flow unsteadiness in simplified Bernoulli's equation significantly affects the pulmonary transvalvular pressure drop estimation. MATERIALS AND METHODS: 3.0T magnetic resonance imaging (MRI) 4D velocity mapping was performed on four healthy volunteers, seven patients with repaired tetralogy of Fallot, and thirteen patients with transposition of the great arteries repaired by arterial switch. Pulmonary transvalvular pressure drop was estimated based on two methods: General Bernoulli's Equation (GBE), ie, the most complete form; and Simplified Bernoulli's Equation (SBE), known as 4V(2) . More than 2300 individual pressure drop measurements were used to compare the simplified and the general Bernoulli's methods. A linear mixed-effects model was employed for statistical analyses, fully accounting for clustering of observations among the methods and systolic phases. RESULTS: The simplified Bernoulli's method systematically underestimated the pressure drop compared to general Bernoulli's method during the entire systolic phase (P < 0.05), including the peak systole, where on average ΔpSBE/ΔpGBE=78%. CONCLUSION: The simplified Bernoulli method underestimated the pressure drop during all systolic phases in all the studied subjects. Therefore, it is necessary to take into account the flow unsteadiness for more accurate estimation of the pressure drop. J. Magn. Reson. Imaging 2016;43:1313-1319.
Building similarity graph...
Analyzing shared references across papers
Loading...
Falahatpisheh et al. (Thu,) conducted a observational in Repaired congenital heart defects and healthy volunteers (n=24). Simplified Bernoulli's Equation (SBE) vs. General Bernoulli's Equation (GBE) was evaluated on Pulmonary transvalvular pressure drop at peak systole (mmHg) (Ratio 0.82, 95% CI 0.700-0.943, p=0.0042). The simplified Bernoulli's method significantly underestimated pulmonary transvalvular pressure drop at peak systole compared to the general Bernoulli's equation (ratio 0.82).
synapsesocial.com/papers/6a1c8e0f66d062ff2dc3f44f — DOI: https://doi.org/10.1002/jmri.25097
Ahmad Falahatpisheh
University of California, Irvine
Carsten Rickers
Deutsches Herzzentrum München
Dominik Daniel Gabbert
Pediatric Cardiology
Journal of Magnetic Resonance Imaging
University of California, Irvine
University of Lübeck
Royal Brompton Hospital
Building similarity graph...
Analyzing shared references across papers
Loading...
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: