Modifying the minimization function in an inverse cardiac generator model reduces dipole crosstalk.
Answers are grounded in this paper and cite their sources.
May improve inverse cardiac model accuracy; leaves open clinical validation before practice change.
Previous work has demonstrated that dipole crosstalk occurs in an inverse cardiac generator consisting of multiple fixed dipoles when only the summed-square error of the potentials is minimized. This paper provides evidence that when cardiac electrical activity can be approximated by a single locatable dipole, crosstalk can be decreased by adding an additional term to the function that is minimized.
No takes yet. Share an insight, caveat, or question.
Ideker et al. (1978) studied this question.