Why the study?
Consistent R-wave amplitude is essential for optimal arrhythmia detection in insertable cardiac monitors, but its variation across postures and activities over time required assessment.
Does insertable cardiac monitor orientation and time since insertion affect R-wave amplitude stability across postures in patients requiring arrhythmia monitoring?
Does insertable cardiac monitor orientation and time since insertion affect R-wave amplitude stability across postures in patients requiring arrhythmia monitoring?
Answers are grounded in this paper and cite their sources.
Inserting cardiac monitors at a 45° angle to the sternum yields higher R-wave amplitudes, which improve and stabilize across different postures by 30 days post-insertion.
BackgroundInsertable cardiac monitors (ICMs) are accepted tools in cardiac arrhythmia management. Consistent R-wave amplitude (RWA) is essential for optimal detection.ObjectivesAssess RWAs with posture/activities at insertion and at 30 days.MethodsParticipants (n = 90) with Confirm Rx™ ICM had RWAs measured in different postures (supine, right-side [RS], left-side [LS], sitting, and standing) and defined physical activities (including isometric push [IPUSH] and pull) at 2 time points. ICMs were inserted in 45° to sternum and parasternal orientations.ResultsThere were significant reductions at insertion with RS, LS, sitting, or standing vs supine (reference position) (all P < .05). At 30 days, significant changes only occurred with LS and sitting (P < .05). Sex had an effect on RWAs, with females having significant variability at insertion (supine vs RS, LS, sitting, standing, and IPUSH; all P < .05). Males showed large RWA interpatient variabilities but minimal differences between positions vs supine. At 30 days, RS, LS, and sitting positions remained significant for females (P < .05), while in males RWAs were higher than at insertion for most postures and activities. The orientation 45° to sternum had consistently higher RWAs vs parasternal orientation at both time points (P < .0001). In females, ICM orientation had no significant effect on RWAs; however, in males the 45° to sternum produced higher RWAs. ICM movement from the insertion site showed no correlation with RWA changes.ConclusionThe mean RWAs were higher at 30 days with less interparticipant and interpostural variability; males had higher RWAs compared to females; 45° to sternum orientation had higher RWAs; and ICM migration from the insertion site did not affect RWAs. Insertable cardiac monitors (ICMs) are accepted tools in cardiac arrhythmia management. Consistent R-wave amplitude (RWA) is essential for optimal detection. Assess RWAs with posture/activities at insertion and at 30 days. Participants (n = 90) with Confirm Rx™ ICM had RWAs measured in different postures (supine, right-side [RS], left-side [LS], sitting, and standing) and defined physical activities (including isometric push [IPUSH] and pull) at 2 time points. ICMs were inserted in 45° to sternum and parasternal orientations. There were significant reductions at insertion with RS, LS, sitting, or standing vs supine (reference position) (all P < .05). At 30 days, significant changes only occurred with LS and sitting (P < .05). Sex had an effect on RWAs, with females having significant variability at insertion (supine vs RS, LS, sitting, standing, and IPUSH; all P < .05). Males showed large RWA interpatient variabilities but minimal differences between positions vs supine. At 30 days, RS, LS, and sitting positions remained significant for females (P < .05), while in males RWAs were higher than at insertion for most postures and activities. The orientation 45° to sternum had consistently higher RWAs vs parasternal orientation at both time points (P < .0001). In females, ICM orientation had no significant effect on RWAs; however, in males the 45° to sternum produced higher RWAs. ICM movement from the insertion site showed no correlation with RWA changes. The mean RWAs were higher at 30 days with less interparticipant and interpostural variability; males had higher RWAs compared to females; 45° to sternum orientation had higher RWAs; and ICM migration from the insertion site did not affect RWAs.
No takes yet. Share an insight, caveat, or question.
Swale et al. (2021) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: