PURPOSE: To compare commercially available 14-guage core biopsy guns and 11- and 14-gauge, directional, vaccum-assisted biopsy probes for tissue yield and fragmentation. MATERIALS AND METHODS: One hundred passes were made through fresh turkey breasts by using each of 11 14-gauge, automated biopsy gun-needle combinations and a directional, stereotactic, vacuum-assisted biopsy system equipped with 11- or 14-gauge probes. Specimens were measured for total weight, individual length, diameter, and number of fragments. RESULTS: Among the handheld devices, the Pro-Mag 2.2 gun (Manan Medical Products, Northbrook, Ill) and 14-gauge Biopsy-Cut needle (Bard, Covington, Ga) had the greatest yield (17.7 mg per core specimen); this combination also produced the greatest percentage of nonfragmented cores (83%). Directional, vacuum-assisted biopsy specimens were significantly larger (P < .001): 36.8 mg per core specimen for the 14-gauge probe and 94.4 mg per core specimen for the 11-gauge probe. Specimens obtained with directional, vacuum-assisted biopsy were more fragmented than those obtained with spring-actuated, two-stage, long-throw, automated biopsy systems. CONCLUSION: Statistically significant differences with various biopsy devices are seen in the breast tissue yield and in fragmentation.
No takes yet. Share an insight, caveat, or question.
Berg et al. (1997) studied this question.