Why the study?
Does high-rate atrial overdrive suppression pacing reduce the incidence of SVT episodes compared to low-rate pacing in patients with medically refractory SVTs?
Does high-rate atrial overdrive suppression pacing reduce the incidence of SVT episodes compared to low-rate pacing in patients with medically refractory SVTs?
Chronic atrial overdrive suppression pacing does not reduce the overall incidence of SVT episodes during prolonged monitoring in patients with medically refractory SVTs.
Answers are grounded in this paper and cite their sources.
High-rate overdrive pacing does not reduce SVT incidence; challenges prior suggestions of efficacy in refractory SVT.
Chronic overdrive suppression pacing has been suggested as an effective adjunctive method for reducing the incidence of cardiac tachyarrhythmias. Documentation of effectiveness during prolonged monitoring is lacking, however. To assess more accurately the long-term utility of this treatment modality for medically refractory supraventricular tachyarrhythmias (SVTs), 10 patients with atrially implanted Intermedics Intertach pacemakers were randomly assigned to either a low or a high bradycardia (back-up) pacing rate. SVT counts were performed during matching follow-up periods both at the initial rate and after rate crossover. The primary antitachycardia modality of this pacemaker (P mod) provides burst pacing to terminate tachycardia episodes, and P mod counters were utilized to quantitate SVT episodes. Tachycardia termination algorithms were programmed to "no restart" and were not changed during the study. The P mod use counter, therefore, reflected the number of discrete episodes of SVTs. Pacemaker implantation diagnoses include atrial flutter, concealed bypass tract, AV nodal reentry, intraatrial reentry, and Wolff-Parkinson-White associated tachycardia. Patient age was 59 +/- 18 yrs. The average pacemaker back-up low rate was 45.7 +/- 4 versus a back-up high rate of 85.1 +/- 2 beats/min. Follow-up was for 57.4 days +/- 33 days at the low rate and 57.3 days +/- 34 days at the high rate (r = 0.99). There was no difference in SVT incidence with a P mod usage of 98.4 +/- 106 at the low rate and 100.8 +/- 94 at the high rate (p = NS). In this blinded, randomized cross-over trial, chronic atrial overdrive suppression pacing did not reduce the overall incidence of SVT episodes during prolonged monitoring.
No takes yet. Share an insight, caveat, or question.
Shandling et al. (1992) studied this question.