Key result
A portable PPG-based cardiovascular monitoring system successfully distinguished between patients with arrhythmia and healthy controls using time, frequency, and nonlinear parameters (p<0.05).
Why the study?
Can a portable PPG sensor system accurately distinguish between patients with arrhythmia and healthy controls?
Observational (n=82)
Can a portable PPG sensor system accurately distinguish between patients with arrhythmia and healthy controls?
p-value: p=<0.05
A novel portable PPG sensor system can accurately record signals and mathematically distinguish between patients with arrhythmia and healthy controls, offering potential for remote cardiovascular monitoring.
Feasibility shown for PPG arrhythmia distinction; hypothesis-generating and requires prospective validation before clinical use.
This article introduces an integrated photoplethysmographic (PPG) based cardiovascular monitoring system that consists of an individually portable PPG device for recording photoplethysmographic signals and a software system with a serverless architecture for processing, storing, and analyzing the obtained signals. The portable device uses the optical plethysmography technique for measuring blood volume in blood vessels. The device was tested and validated by a comparative analysis of three photoplethysmographic signals and one Electrocardiographic signal registered simultaneously in the target subject. The comparative analysis of these signals shows insignificant deviations in the obtained results, with the mean squared error between the studied signals being less than 21 ms. This deviation cannot affect the results that were obtained from the analysis of the interval series tested. Based on this result, we assume that the detected signals with the proposed device are realistic. The designed software system processes the registered data, performs preprocessing, determines the pulse rate variability, and performs mathematical analysis of PP intervals. Two groups of subjects were studied: 42 patients with arrhythmia and 40 healthy controls. Mathematical methods for data analysis in time and frequency domain and nonlinear methods (Poincaré plots, Rescaled Range Plot, Detrended Fluctuation Analysis, and MultiFractal Detrended Fluctuation Analysis) are applied. The obtained results are presented in tabular form and some of them in graphical form. The parameters studied in the time and frequency domain, as well as with the nonlinear methods, have statistical significance (p < 0.05) and they can distinguish between the two studied groups. Visual analysis of PP intervals, based on Poincare’s nonlinear method, provides important information on the physiological status of patients, allowing for one to see at a glance the entire PP interval series and quickly detect cardiovascular disorders, if any. The photoplethysmographic data of healthy individuals and patients diagnosed with arrhythmia were recorded, processed, and examined through the system under the guidance of a cardiologist. The results were analyzed and it was concluded that this system could serve to monitor patients with cardiovascular diseases and, when the condition worsens, a signal could be generated and sent to the hospital for undertaking immediate measures to stabilize patient’s health.
No takes yet. Share an insight, caveat, or question.
Georgieva-Tsaneva et al. (2020) conducted an observational in Arrhythmia (n=82). Portable PPG-based cardiovascular monitoring system vs. Healthy controls was evaluated on Differentiation between arrhythmia patients and healthy controls using time, frequency, and nonlinear parameters (p=<0.05). A portable PPG-based cardiovascular monitoring system successfully distinguished between patients with arrhythmia and healthy controls using time, frequency, and nonlinear parameters (p<0.05).