Key result
A novel mapping algorithm integrating vector data identified the actual mechanism of scar-related atrial tachycardias more frequently than the standard algorithm (67.2% vs. 44.8%, p=0.009).
Why the study?
Activation maps of scar-related AT can be challenging to interpret due to inaccurate electrogram annotation and an arbitrarily predefined mapping window, prompting evaluation of a novel mapping algorithm.
Does a novel mapping algorithm integrating vector data improve the identification of the mechanism of scar-related atrial tachycardias compared to standard algorithms?
Observational (n=77)
Does a novel mapping algorithm integrating vector data improve the identification of the mechanism of scar-related atrial tachycardias compared to standard algorithms?
A novel mapping algorithm integrating vector data significantly improves the identification of arrhythmia mechanisms in scar-related atrial tachycardias compared to standard mapping.
May aid mechanism identification in scar-related atrial tachycardias; hypothesis-generating and should not yet change practice.
BACKGROUND: Activation maps of scar-related atrial tachycardias (AT) can be challenging to interpret due to difficulty in inaccurate annotation of electrograms, and an arbitrarily predefined mapping window. A novel mapping software integrating vector data and applying an algorithmic solution taking into consideration global activation pattern has been recently described (Coherent™, Biosense Webster "Investigational"). OBJECTIVE: We aimed to assess the investigational algorithm to determine the mechanism of AT compared with the standard algorithm. METHODS: This study included patients who underwent ablation of scar-related AT using the Carto 3 and the standard activation algorithm. The mapping data were analyzed retrospectively using the investigational algorithm, and the mechanisms were evaluated by two independent electrophysiologists. RESULTS: A total of 77 scar-related AT activation maps were analyzed (89.6% left atrium, median tachycardia cycle length of 273 ms). Of those, 67 cases with a confirmed mechanism of arrhythmia were used to compare the activation software. The actual mechanism of the arrhythmia was more likely to be identified with the investigational algorithm (67.2% vs. 44.8%, p = .009). In five patients with dual-loop circuits, 3/5 (60%) were correctly identified by the investigational algorithm compared to 0/5 (0%) with the standard software. The reduced atrial voltage was prone to lead to less capable identification of mechanism (p for trend: .05). The investigational algorithm showed higher inter-reviewer agreement (Cohen's kappa .62 vs. .47). CONCLUSIONS: In patients with scar-related ATs, activation mapping algorithms integrating vector data and "best-fit" propagation solution may help in identifying the mechanism and the successful site of termination.
No takes yet. Share an insight, caveat, or question.
Kuroda et al. (2021) conducted an observational in Scar-related atrial tachycardias (n=77). Investigational mapping algorithm (Coherent) vs. Standard activation algorithm was evaluated on Identification of the actual mechanism of the arrhythmia (p=0.009). A novel mapping algorithm integrating vector data identified the actual mechanism of scar-related atrial tachycardias more frequently than the standard algorithm (67.2% vs. 44.8%, p=0.009).