Key result
Auto2DE was 6.5-7.5 fold faster than manual 2DE and showed excellent agreement in healthy mice with good image quality, but performed poorly in diabetic animal models.
Why the study?
Manual tracing of endocardial borders in small animal echocardiography is time consuming and varies with operator experience.
Does an automated 2D-border detection algorithm (Auto2DE) improve the speed and maintain accuracy of LV function assessment compared to manual 2DE in small animal models?
Does an automated 2D-border detection algorithm (Auto2DE) improve the speed and maintain accuracy of LV function assessment compared to manual 2DE in small animal models?
Answers are grounded in this paper and cite their sources.
Accelerates preclinical LV assessment in healthy models; leaves open reliability in disease states and clinical translation.
Echocardiography is the most commonly applied technique for non-invasive assessment of cardiac function in small animals. Manual tracing of endocardial borders is time consuming and varies with operator experience. Therefore, we aimed to evaluate a novel automated two-dimensional software algorithm (Auto2DE) for small animals and compare it to the standard use of manual 2D-echocardiographic assessment (2DE). We hypothesized that novel Auto2DE will provide rapid and robust data sets, which are in agreement with manually assessed data of animals.2DE and Auto2DE were carried out using a high-resolution imaging-system for small animals. First, validation cohorts of mouse and rat cine loops were used to compare Auto2DE against 2DE. These data were stratified for image quality by a blinded expert in small animal imaging. Second, we evaluated 2DE and Auto2DE in four mouse models and four rat models with different cardiac pathologies.Automated assessment of LV function by 2DE was faster than conventional 2DE analysis and independent of operator experience levels. The accuracy of Auto2DE-assessed data in healthy mice was dependent on cine loop quality, with excellent agreement between Auto2DE and 2DE in cine loops with adequate quality. Auto2DE allowed for valid detection of impaired cardiac function in animal models with pronounced cardiac phenotypes, but yielded poor performance in diabetic animal models independent of image quality.Auto2DE represents a novel automated analysis tool for rapid assessment of LV function, which is suitable for data acquisition in studies with good and very good echocardiographic image quality, but presents systematic problems in specific pathologies.
No takes yet. Share an insight, caveat, or question.
Grune et al. (2019) studied Healthy and diseased small animals (mice and rats). Auto2DE (automated 2D-border detection algorithm) vs. Conventional manual 2D-echocardiography (2DE) was evaluated on Correlation and agreement of LV function parameters between Auto2DE and 2DE. Auto2DE was 6.5-7.5 fold faster than manual 2DE and showed excellent agreement in healthy mice with good image quality, but performed poorly in diabetic animal models.
Synapse has enriched 10 related papers. Consider them for comparative context: