A new 14 C detector system containing nine, independently working, CO 2 proportional counters is described. The system is designed for a sufficient measuring capacity at a precision level better than σ = ± 2‰, which requires a counting time of about one week per sample. The size of the installation requires a simple and economic design of counters and electronics. A single anticoincidence shield for all counters consists of five newly developed flat counters. The modern counting rate (52cpm) is sensitively checked by running Heidelberg sodium carbonate standard samples wth a counting rate of about 10 times modern. A microcomputer (DEC PDP-11 /03) is used for data acquisition. Recent developments in laboratory techniques (preparation and gaschromatographic purification of samples) are also reported.