This is the authors' abstract. We don't add key points for this paper.
Background and objective: Virtual reality (VR) simulators enrich surgical training and offer training possibilities outside of the operating room (OR). In this study, we created a criterion-based training program on a VR simulator with haptic feedback and tested it by comparing the performances of a simulator group against a control group.Material and methods: Medical students with no experience in laparoscopy were randomly assigned to a simulator group or a control group. In the simulator group, the candidates trained until they reached predefined criteria on the LapSim® VR simulator (Surgical Science AB, Göteborg, Sweden) with haptic feedback (XitactTM IHP, Mentice AB, Göteborg, Sweden). All candidates performed a cholecystectomy on a porcine organ model in a box trainer (the clinical setting). The performances were video rated by two surgeons blinded to subject training status.Results: In total, 30 students performed the cholecystectomy and had their videos rated (N = 16 simulator group, N = 14 control group). The control group achieved better video rating scores than the simulator group (p < .05).Conclusions: The criterion-based training program did not transfer skills to the clinical setting. Poor mechanical performance of the simulated haptic feedback is believed to have resulted in a negative training effect.
Answers are grounded in this paper and cite their sources.
No takes yet. Share an insight, caveat, or question.
Våpenstad et al. (2017) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: