Why the study?
What is the incidence of and factors predicting skin burns at the indifferent electrode site during radiofrequency catheter ablation?
What is the incidence of and factors predicting skin burns at the indifferent electrode site during radiofrequency catheter ablation?
Skin burns at the indifferent electrode site during RFA are a rare complication (0.28%) associated with higher BMI and longer procedure times.
Skin burns during RFA remain rare yet associate with higher BMI and longer procedures; leaves open targeted prevention trials.
Radiofrequency catheter ablation (RFA) has become a mainstay for treatment of cardiac arrhythmias. Skin burns at the site of an indifferent electrode patch have been a rare, serious, and likely an underreported complication of RFA. The purpose of this study was to determine the incidence of skin burns in cardiac RFA procedures performed at one institution. Also, we wanted to determine the factors predicting skin burns after cardiac RFA procedures at the indifferent electrode skin pad site. Methods. A retrospective case control study was performed to compare the characteristics in patients who developed skin burns in a 2-year period. Results. Incidence of significant skin burns after RFA was 0.28% (6/2167). Four of the six patients were female and all were Caucasians. Four controls for every case were age and sex matched. Burn patients had significantly higher BMI, procedure time, and postprocedure pain, relative to control subjects (p < 0.05, one-tailed testing). No one in either group had evidence of dispersive pad malattachment. Conclusions. Our results indicate that burn patients had higher BMI and longer procedure times compared to control subjects. These findings warrant further larger studies on this topic.
No takes yet. Share an insight, caveat, or question.
Ibrahim et al. (2016) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: