PURPOSE: To compare short tau inversion recovery (STIR) images with slice encoding for metal artifact correction (SEMAC)-corrected magnetic resonance imaging (MRI) of spectral presaturation with inversion recovery (SPIR) or inversion recovery (IR) at 3T in patients with metallic spinal instrumentation. Two musculoskeletal radiologists compared the image sets and qualitatively analyzed the images with a five-point scale that included artifact reduction around the metallic implant and visualization of the rod and pedicle. Quantitative assessments were performed by calculating the signal intensity ratio of the fixated vertebra and non-metallic vertebra and by calculating the signal-to-noise ratios (SNRs) of the vertebrae. A paired t-test was used for the statistical analyses. RESULTS: The SEMAC-IR MRI had a significant decrease in the metallic artifact area (P < 0.05), while the SEMAC-SPIR MRI yielded significantly increased artifact areas (P < 0.05). However, the signal intensity ratios (i.e., metal-induced signal pile-up) were not significantly different (P > 0.05) between the STIR and SEMAC MRI. The SNR of the SEMAC MRI was significantly lower than the SNR of the STIR (P < 0.05). The metal artifact reduction scores were significantly higher in the SEMAC-SPIR MRI (P < 0.05). However, the drawbacks of high signal pile-up, large artifact size, and relatively low SNRs require further investigation to determine the best method for fat-suppressed MRI of metallic implants.
No takes yet. Share an insight, caveat, or question.
Lee et al. (2016) studied this question.