Answers are grounded in this paper and cite their sources.
Intraoperative ultrasonography (IoUSG) is an essential real-time imaging tool in neurosurgical practice. However, a physical mismatch between standard probe dimensions and narrow craniotomy windows often hinders its effectiveness. We present a practical and cost-effective "balloon technique" using a saline-filled sterile surgical glove to establish an acoustic bridge. A 61-year-old male with a metastatic brain lesion in the right frontal region underwent a 3x3.5 cm crani-otomy, where a standard 4 cm wide probe exceeded the bone margins. By utilizing a saline-filled interface, we successful-ly achieved high-resolution visualization and total resection without enlarging the craniotomy. This technique provides real-time feedback and brain-shift compensation while maintaining the principles of minimally invasive surgery, offering a superior alternative to costly and time-consuming intraoperative imaging modalities.
No takes yet. Share an insight, caveat, or question.
Özmarasalı et al. (2026) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: