Key result
Non-invasive oscillometric blood pressure measurement overestimated intra-arterial blood pressure, with a mean difference of -3.6 mm Hg for systolic blood pressure.
Why the study?
Intra-arterial blood pressure measurement is the gold standard in critically ill children but carries risks and lacks widespread availability, prompting comparison with oscillometric non-invasive measurement.
Does oscillometric non-invasive blood pressure measurement accurately reflect intra-arterial blood pressure in critically ill children in the PICU?
Observational (n=65)
No
Does oscillometric non-invasive blood pressure measurement accurately reflect intra-arterial blood pressure in critically ill children in the PICU?
Answers are grounded in this paper and cite their sources.
Non-invasive blood pressure measurement significantly overestimates blood pressure compared to invasive arterial monitoring in critically ill children, particularly during hypotension and in infants.
Intra-arterial blood pressure (IABP) measurement, although considered the gold standard in critically ill children, is associated with certain risks and lacks widespread availability. This study was conducted to determine the differences and agreements between oscillometric non-invasive blood pressure (NIBP) and invasive IABP measurements in children. Inclusion criteria consisted of children (from 1 month to 18 years) admitted to the pediatric intensive care unit (PICU) of a teaching hospital who required arterial catheter insertion for blood pressure (BP) monitoring. The comparison between IABP and NIBP was studied using paired t-test, Bland–Altman analysis, and Pearson's correlation coefficient. In total, 4,447 pairs of simultaneously recorded hourly NIBP and IABP measurements were collected from 65 children. Mean differences between IABP and NIBP were −3.6 ± 12.85, −4.7 ± 9.3, and −3.12 ± 9.30 mm Hg for systolic, diastolic, and mean arterial BP, respectively (p < 0.001), with wide limits of agreement. NIBP significantly overestimated BP (p < 0.001) in all three BP states (hypotensive, normotensive, and hypertensive), except systolic blood pressure (SBP) during hypertension where IABP was significantly higher. The difference in SBP was most pronounced during hypotension. The difference in SBP was significant in children <10 years (p < 0.001), with the maximum difference being in infants. It was insignificant in adolescents (p = 0.28) and underweight children (p = 0.55). NIBP recorded significantly higher BP in all states of BP except SBP in the hypertensive state. SBP measured by NIBP tended to be the most reliable in adolescents and underweight children. NIBP was the most unreliable in infants, obese children, and during hypotension.
No takes yet. Share an insight, caveat, or question.
Kaur et al. (2021) conducted an observational in Critically ill children requiring blood pressure monitoring (n=65). Non-invasive oscillometric blood pressure (NIBP) measurement vs. Intra-arterial blood pressure (IABP) measurement was evaluated on Mean difference between IABP and NIBP for systolic blood pressure (MD -3.6 mm Hg, p=<0.001). Non-invasive oscillometric blood pressure measurement overestimated intra-arterial blood pressure, with a mean difference of -3.6 mm Hg for systolic blood pressure.