Key result
Continuous noninvasive finger blood pressure measurement showed an overall mean bias of 0.10 mm Hg compared to invasive monitoring, with 95% of measurements differing by ≤10 mm Hg.
Why the study?
Does continuous noninvasive finger blood pressure measurement accurately reflect invasive arterial blood pressure in critically ill patients?
Observational (n=39)
No
Does continuous noninvasive finger blood pressure measurement accurately reflect invasive arterial blood pressure in critically ill patients?
Noninvasive continuous finger blood pressure measurement using a finger cuff is generally accurate in critically ill patients, though significant discrepancies can occasionally occur.
Answers are grounded in this paper and cite their sources.
May support finger-cuff monitoring when arterial access unavailable; leaves open prospective validation before routine adoption.
No abstract yet. The journal hasn't published one we can show.
No takes yet. Share an insight, caveat, or question.
Hirschl et al. (1996) conducted an observational in Critically ill requiring invasive arterial blood pressure monitoring (n=39). Continuous noninvasive finger blood pressure measurement vs. Invasive mean arterial blood pressure measurement (radial artery cannulation) was evaluated on Discrepancy between noninvasive finger cuff and intra-arterial blood pressure measurements (Mean discrepancy 0.10 mm Hg). Continuous noninvasive finger blood pressure measurement showed an overall mean bias of 0.10 mm Hg compared to invasive monitoring, with 95% of measurements differing by ≤10 mm Hg.