The conductance catheter system provides accurate in vivo measurements of murine left ventricular volumes, validating its use for hemodynamic research.
The purpose of this study was to define the validation methods and outcomes of a conductance catheter system specifically for in vivo murine cardiac hemodynamic analysis. To express the relationship between conductance and blood volumes, we used an in vitro model to derive a volume-conductance line. The volume-conductance line was used to compute raw volume from the modified conductance signals. The parallel volume was calibrated with hypertonic (15%) saline injected from extrajugular vein. The ventricular volume was computed by raw volume minus parallel volume. The accuracy of conductance volumetric measurements was validated with a static in situ infusion of calibrated volumes of whole blood injected into arrested left ventricles. In vivo dynamic measurements were performed with 24 C57B1/6 mice, 6 months old; for comparison of established values. The in situ model showed that after calibration, the experimental coefficient, alpha, was equal to 1 and the measured volume by conductance catheter was equal to the true volume of the left ventricle (y = 0.982x + 0.513, p < .0001). For the in vivo models, the end-diastolic volumes and the stroke volumes and cardiac output determined with the conductance catheter system were 17.3 +/- 1.0 microL, 10.6 +/- 0.9 microL, and 6.0 +/- 0.5 mL/min, respectively. We validated the relationship between measured volume by conductance catheter and the true volume and demonstrated the accuracy of the volume-conductance line for conversion of conductance to volume.