An apparatus is described which measures the transmission of a beam of 60 keV γ radiation from 241Am as it is scanned across a selected site in a patient. The transmitted count-rate is either displayed logarithmically for visual inspection of bone thickness profiles or collected on paper tape for analysis by computer. Various parameters to describe bone density are denned and discussed with particular reference to studies of the femur. The primary “bone index” derived from a scan represents a mass of bone per unit length; it is particularly useful in serial measurements and has a reproducibility of about 2 per cent. For inter-patient comparisons an “effective bone density” (defined as bone index ÷ cross-sectional area) is preferred; it shows a coefficient of variation of about 3 per cent. The accuracy and reproducibility of these and other related parameters are discussed. The method enables account to be taken of variations in the soft tissue attenuation consequent upon variability in the muscle/fat ratio. Some clinical applications (which are being reported in full elsewhere) are briefly noted.