Key result
A deep neural network using an optimal feature set from photoplethysmography signals accurately estimated continuous blood pressure, achieving a mean absolute error of 2.480 mmHg for systolic and 1.499 mmHg for diastolic blood pressure.
Why the study?
Blood pressure is currently monitored using uncomfortable cuff-based devices that do not support continuous monitoring, necessitating noninvasive methods based on photoplethysmography.
Does a deep neural network algorithm using an optimal feature set from PPG signals accurately estimate continuous blood pressure?
Observational (n=219)
Does a deep neural network algorithm using an optimal feature set from PPG signals accurately estimate continuous blood pressure?
Effect estimate: MAE 2.480 mmHg (SBP) and 1.499 mmHg (DBP)
A novel deep learning algorithm using an optimal feature set from PPG signals can accurately estimate continuous systolic and diastolic blood pressure, meeting clinical standards.
Supports PPG-based cuffless BP monitoring research; leaves open prospective validation before clinical adoption.
Continuous Blood Pressure (BP) provides essential information for monitoring one's health condition. However, BP is currently monitoring using uncomfortable cuff-based devices, which does not support continuous BP monitoring. This paper aims to introduce a blood pressure monitoring algorithm based on only photoplethysmography (PPG) signals using the deep neural network (DNN). The PPG signals are obtained from 219 subjects with 657 records and filtered using signal processing algorithms to reduce the effects of noise, such as baseline wandering, and motion artifacts. The proposed algorithm is based on pulse wave analysis of PPG signals, extracted various domain features from PPG signals, and mapped them to BP values. Four feature selection methods are applied and yielded four feature subsets. Therefore, an ensemble feature selection technique is proposed to obtain the optimal feature set based on major voting scores from four feature subsets. DNN models, along with the ensemble feature selection technique, outperformed in estimating the systolic blood pressure (SBP) and diastolic blood pressure (DBP) compared to previously reported approaches that rely only on the PPG signal. The coefficient of determination (R2) and mean absolute error (MAE) of the proposed algorithm are 0.962 and 2.480 mmHg respectively for SBP, 0.955 and 1.499 mmHg respectively for DBP. The proposed approach meets the Advancement of Medical Instrumentation standard for SBP and DBP estimations. Additionally, according to the British Hypertension Society standard, the results attained Grade A for both SBP and DBP estimations. It concludes that BP can be estimated more accurately using the optimal feature set and DNN models. The proposed algorithm has the potential ability to facilitate mobile healthcare devices to monitor continuous BP.
No takes yet. Share an insight, caveat, or question.
Raju et al. (2023) conducted an observational in Blood pressure estimation (n=219). Deep neural network with ensemble feature selection vs. Reference blood pressure measurements was evaluated on Mean absolute error (MAE) for systolic and diastolic blood pressure estimation (MAE 2.480 mmHg (SBP) and 1.499 mmHg (DBP)). A deep neural network using an optimal feature set from photoplethysmography signals accurately estimated continuous blood pressure, achieving a mean absolute error of 2.480 mmHg for systolic and 1.499 mmHg for diastolic blood pressure.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: