Answers are grounded in this paper and cite their sources.
Background: Membrane potential ( V mem ) exerts physiological influence across a wide range of time and space scales. To study V mem in these diverse contexts, it is essential to accurately record absolute values of V mem , rather than solely relative measurements. Materials and Methods: We use fluorescence lifetime imaging of a small molecule voltage sensitive dye ( VF2.1.Cl ) to estimate mV values of absolute membrane potential. Results: We test the consistency of VF2.1.Cl lifetime measurements performed on different single-photon counting instruments and find that they are in striking agreement (differences of <0.5 ps/mV in the slope and <50 ps in the y -intercept). We also demonstrate that VF2.1.Cl lifetime reports absolute V mem under two-photon (2P) illumination with better than 20 mV of V mem resolution, a nearly 10-fold improvement over other lifetime-based methods. Conclusions: We demonstrate that VF-FLIM is a robust and portable metric for V mem across imaging platforms and under both one-photon and 2P illumination. This work is a critical foundation for application of VF-FLIM to record absolute membrane potential signals in thick tissue.
No takes yet. Share an insight, caveat, or question.
Lazzari-Dean et al. (2021) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: