Does a novel 1.5T MRI-conditional 12-lead ECG system provide high-fidelity ECG monitoring during MRI?
Does a novel 1.5T MRI-conditional 12-lead ECG system provide high-fidelity ECG monitoring during MRI?
Answers are grounded in this paper and cite their sources.
A novel 1.5T MRI-conditional 12-lead ECG system successfully provides high-fidelity intra-MRI ECG monitoring with minimal noise and heating.
PURPOSE: High-fidelity 12-lead electrocardiogram (ECG) is important for physiological monitoring of patients during MR-guided intervention and cardiac MRI. Issues in obtaining noncorrupted ECGs inside MRI include a superimposed magneto-hydro-dynamic voltage, gradient switching-induced voltages, and radiofrequency heating. These problems increase with magnetic field. The aim of this study is to develop and clinically validate a 1.5T MRI-conditional 12-lead ECG system. METHODS: The system was constructed with transmission lines to reduce radiofrequency induction and switching circuits to remove induced voltages. Adaptive filters, trained by 12-lead measurements outside MRI and in two orientations inside MRI, were used to remove the magneto-hydro-dynamic voltage. The system was tested on 10 (one exercising) volunteers and four arrhythmia patients. RESULTS: Switching circuits removed most imaging-induced voltages (residual noise <3% of the R-wave). Magneto-hydro-dynamic voltage removal provided intra-MRI ECGs that varied by <3.8% from those outside the MRI, preserving the true S-wave to T-wave segment. In premature ventricular contraction (PVC) patients, clean ECGs separated premature ventricular contraction and sinus rhythm beats. Measured heating was <1.5°C. The system reliably acquired multiphase (steady-state free precession) wall-motion-cine and phase-contrast-cine scans, including subjects in whom 4-lead gating failed. The system required a minimum repetition time of 4 ms to allow robust ECG processing. CONCLUSION: High-fidelity intra-MRI 12-lead ECG is possible.
No takes yet. Share an insight, caveat, or question.
Tse et al. (2013) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: