Key result
Transcutaneous range gated Doppler velocimeter measurements of femoral blood flow showed a highly significant correlation with electromagnetic probe measurements.
Answers are grounded in this paper and cite their sources.
Supports range-gated Doppler development in vitro; leaves open clinical translation for femoral flow assessment.
Femoral blood flow (FBF) was measured in seven dogs, simultaneously with both an electromagnetic perivascular probe and a transcutaneous range gated Doppler velocimeter. Measurements were made in basal conditions and during intraarterial infusions of noradrenaline (10 to 400 ng . kg-1 . min-1) and isoprenaline (10 to 400 ng . kg-1 . min-1) thus allowing comparisons of ultrasonic (DBF) and electromagnetic (EMBF) blood flow at 91 different blood flow rates ranging from 5 to 300 cm3 . min-1. The linear regression line through the data of ultrasonic and electromagnetic simultaneous measurement was: DBF = 0.8 + 1.016 EMBF +/- 19.0 cm3 . min-1 with a highly significant correlation (r = 0.96, p less than 0.001) but there was a wide scattering about the mean. Errors in DBF were mainly due to positioning of the probe and determination of arterial diameter.
No takes yet. Share an insight, caveat, or question.
Chauveau et al. (1985) studied this question. Transcutaneous range gated Doppler velocimeter vs. Electromagnetic perivascular probe was evaluated on Correlation between ultrasonic and electromagnetic blood flow measurements (r = 0.96, p=<0.001). Transcutaneous range gated Doppler velocimeter measurements of femoral blood flow showed a highly significant correlation with electromagnetic probe measurements.