A high pressure photoemission system is described that combines differential pumping with an electrostatic lens system. This approach allows optimized differential pumping without loss of signal, thereby increasing the high-pressure performance by at least 2 orders of magnitude compared to passive differential pumping systems. A general analysis of aperture-based high-pressure photoemission is presented, followed by a description of the prototype system which has operated at pressures up to 7 mbar on a synchrotron beamline. Using this approach, photoemission experiments should be possible up to 100 mbar. Example data are presented for dielectric samples in gas atmospheres, for a copper catalyst under reaction conditions, and for liquid water in equilibrium with its vapor.
No takes yet. Share an insight, caveat, or question.
Ogletree et al. (2002) studied this question.
Synapse has enriched one closely related paper. Consider it for comparative context: