Key result
The EEMD-based R-peak detection algorithm achieved a positive predicted value of 91.08% and sensitivity of 92.76% on low-SNR ECGs acquired during sport activity.
Why the study?
Automatic ECG examination during sport activity is challenged by low signal to noise ratio, complicating the initial identification of R peaks needed for heart rate estimation.
Does an EEMD-based algorithm accurately detect R-peaks in low-SNR ECGs acquired during sport activity?
Does an EEMD-based algorithm accurately detect R-peaks in low-SNR ECGs acquired during sport activity?
An EEMD-based algorithm effectively detects R-peaks and computes heart rate from low-SNR ECGs acquired by portable sensors during physical activity.
Wearable and portable electrocardiographic devices are revolutionizing athlete's screening through digital health application enabling a continuous monitoring of important cardiac parameters in real-time. Automatic examination of electrocardiogram (ECG) acquired during sport activity is challenging because acquisition conditions often lead to record ECGs with low signal to noise ratio (SNR). The initial issue of automatic ECG analysis is the identification of R peaks. R peaks are fundamental for the estimation of heart rate, which is the primary clinical parameter used by athletes for athletic performance evaluation. Thus, the aim of this research is to propose an R-peak detection algorithm for ECGs acquired during sport activity by portable and wearable sensors dealing with low SNR. The algorithm is based on a noise assisted data analysis method: Ensemble Empirical Mode Decomposition method (EEMD). Localization of R peaks is primarily performed on the first intrinsic mode function extracted by the EEMD. The algorithm was tested on `Run on indoor treadmill' dataset from Physionet. ECGs were acquired during running/light jogging on an indoor treadmill and present a low SNR (1±7 dB). The developed EEMD-based algorithm showed good performances in terms of positive predicted value (91.08%), sensitivity (92.76%), false discovery rate (8.92), false negative rate (7.24%), cumulative statistical index (83.84%) and mean R-peak position error 1.10 [0.46;1.46]ms. EEMD-based algorithm performs efficiently also in computing heart rate. In conclusion, the developed R-peak detection EEMD-based algorithm showed good level of performances even working on low-SNR ECG acquired during sport activity by portable sensors.
No takes yet. Share an insight, caveat, or question.
Romagnoli et al. (2021) studied ECGs acquired during sport activity. Ensemble Empirical Mode Decomposition (EEMD) based R-peak detection algorithm was evaluated on R-peak detection performance (positive predicted value and sensitivity). The EEMD-based R-peak detection algorithm achieved a positive predicted value of 91.08% and sensitivity of 92.76% on low-SNR ECGs acquired during sport activity.