Why the study?
Is increased electrode impedance a reliable predictor of missed QRS detections and the need to replace degraded electrocardiographic electrodes?
Is increased electrode impedance a reliable predictor of missed QRS detections and the need to replace degraded electrocardiographic electrodes?
Answers are grounded in this paper and cite their sources.
Increased electrode impedance is not a reliable predictor of a poor ECG electrode and the need to replace it.
We degraded electrocardiographic electrodes by exposing them to air for four days and evaluated them on 12 subjects. After application, we recorded the electrocardiogram (including motion artifact), missed QRS detections and electrode impedance during 5 min of arm and body movements. Missed QRS detections increased with electrode impedance but correlation was poor. Increased electrode impedance was not a reliable predictor of a poor electrode and the need to replace it.
No takes yet. Share an insight, caveat, or question.
Wiese et al. (2004) studied this question.