Answers are grounded in this paper and cite their sources.
We developed a universal liquid chromatography-mass spectrometry (LC-MS) assay with automated online extraction to quantify simultaneously the immunosuppressants sirolimus, everolimus, tacrolimus, and cyclosporine. Whole blood (300 microl) plus 6 microl 32-desmethoxyrapamycin (1 ng/microl) as internal standard was treated with 600 microl methanol/0.2 M ZnSO4 (80/20 v/v). After vortexing (30 s) and centrifugation (20000 g, 5 min) 50 microl of the supernatant were loaded on an extraction column, were washed by 0.35 ml/min water for 3 min and, after activation of a column-switching valve, were back-flushed by 0.25 ml/min methanol/ water (90/10 v/v) onto a C18 analytical column. After 22 min the extraction column was washed for 2 min with methanol and for 3 min with water before starting the next run. Column temperatures were kept at 33 degrees C. Sodium adduct ions [M+Na]+ ions were detected in the selected ion mode. For sirolimus, everolimus and tacrolimus the assay was linear from 0.3 to 200 microg/l and for cyclosporine from 5 to 1000 microg/l (all r2>0.999). Recovery of all immunosuppressants and the internal standard was >90% and in general, inter-day and intraday precision was <10%. The simultaneous quantification of blood levels by LC-MS seems to be the method of choice in transplanted patients receiving multiple immunosuppressants.
No takes yet. Share an insight, caveat, or question.
Deters et al. (2002) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: