Answers are grounded in this paper and cite their sources.
OBJECTIVES: To investigate risk factors for inferior vena cava (IVC) filter abutment, including external compression on the IVC wall, using venous phase computed tomography (CT). METHODS: One-hundred-forty-one cases of Celect IVC filter insertion between January 2009 and April 2017 were retrospectively reviewed. On pre-procedural CT, IVC diameter and morphological classifications were measured. Filter tilt angle, IVC angle, vertical position, and filter tip abutment to the IVC wall were analyzed on post-procedural CT. IVC compression was examined by pre- and post-procedural CT analysis. Multiple logistic regression analysis was conducted to find factors related to IVC filter abutment. In multiple logistic regression analysis, filter-tilt angle over 9.25° and external compression on the IVC were found to be independent predictors of filter abutment (odds ratios: 4.56, 10.18, respectively). CONCLUSION: IVC filter tilt, external compression on IVC wall, and IVC morphology were significantly different between the filter tip abutment and non-abutment groups. External compression and filter tilt over 9.25° were risk factors for filter tip abutment in multiple logistic regression analysis. By identifying these factors, we may be able to reduce filter tilting by preventing the filter from being deployed in a dangerous area.
No takes yet. Share an insight, caveat, or question.
Lee et al. (2018) studied this question.