This theoretical paper clarifies the energy loss approach for analyzing artificial heart valves, providing a derived energy equation and demonstrating its practical suitability.
Answers are grounded in this paper and cite their sources.
Enhances theoretical modeling of valve energy loss; leaves open clinical translation pending validation studies.
The purpose of this paper is to clarify certain aspects of the 'energy loss' approach to the analysis of artificial heart valves. This involves the discussion of what is meant by 'energy loss'; the derivation of an appropriate energy equation; the scrutiny of the conditions under which it may be applied and of the assumptions involved in deriving a useful formulation in terms of readily measurable parameters; consideration of the extent to which these assumptions are valid in reported experiments using this formulation and the demonstration of the suitability and versatility of this approach when applied specifically to heart valve conduits.
No takes yet. Share an insight, caveat, or question.
Leefe et al. (1987) studied this question.