Key result
Leo wearable patch strongly correlates with gold-standard HR and RR measurements in children.
Why the study?
Remote patient monitoring enables virtual care, prompting the development and performance testing of a novel wearable chest patch system (Leo).
Does the 'Leo' wearable remote patient monitoring system accurately measure heart rate and respiration rate compared to gold-standard devices in children?
Observational (n=35)
Does the 'Leo' wearable remote patient monitoring system accurately measure heart rate and respiration rate compared to gold-standard devices in children?
The 'Leo' wearable chest patch demonstrates high accuracy for measuring heart rate and respiration rate in children, supporting its potential clinical utility for remote patient monitoring.
Answers are grounded in this paper and cite their sources.
Supports pediatric remote monitoring feasibility; hypothesis-generating for clinical outcomes in prospective trials.
Remote patient monitoring (RPM) enables healthcare professionals to virtually care for patients. This paper presents a novel RPM system, 'Leo' that includes a wearable chest patch, adhesives, charging cable, mobile application and a cloud-based analytics platform. It records ECG, impedance, accelerometer and gyroscope signals and outputs heart rate (HR), respiration rate (RR) and body position. To test the performance of Leo, 35 children were recruited. The study consisted of 2 clinic visits on consecutive days. During which the Leo system was compared with gold-standard devices for HR and RR while participants performed different breathing maneuvers in different positions. Between visits, participants wore the Leo device for up to 24 hours at-home while continuing their daily activities. During clinic visits, HR measured by Leo was highly correlated with gold-standard (r = 0.99). RR measured by Leo was also very high correlated with gold-standard (r = 0.97). For 24 hour at-home monitoring, the Leo was able to collect on average 17.4 hours of data per participant. These results highlight the potential clinical use of Leo as an RPM system.
No takes yet. Share an insight, caveat, or question.
Javed et al. (2024) reported an observational. Leo remote patient monitoring system vs. Gold-standard devices was evaluated on Heart rate (HR) and respiration rate (RR) correlation with gold-standard devices (r = 0.99 for HR; r = 0.97 for RR). The Leo wearable chest patch demonstrated high correlation with gold-standard devices for measuring heart rate and respiration rate in children.
Synapse has enriched 8 closely related papers on similar clinical questions. Consider them for comparative context: