Key result
Universal LMWH-calibrated anti-Xa assay correlates near-perfectly with DOAC drug levels with ~96% sensitivity.
Why the study?
A universal anti-Xa assay to determine rivaroxaban, apixaban, and edoxaban drug concentrations could simplify laboratory procedures and facilitate widespread implementation.
Does a universal LMWH-calibrated anti-Xa assay accurately measure rivaroxaban, apixaban, and edoxaban concentrations compared to LC-MS/MS in patients treated with these drugs?
Cross-Sectional (n=867)
Blinded to reference standard
Yes
Does a universal LMWH-calibrated anti-Xa assay accurately measure rivaroxaban, apixaban, and edoxaban concentrations compared to LC-MS/MS in patients treated with these drugs?
Effect estimate: rs = 0.98 (95% CI 0.98-0.98)
A universal LMWH-calibrated anti-Xa assay accurately measures DOAC concentrations, simplifying laboratory procedures for clinical practice.
May simplify DOAC level assessment in labs; leaves open prospective validation before clinical adoption.
Summary A universal anti‐Xa assay for the determination of rivaroxaban, apixaban and edoxaban drug concentrations would simplify laboratory procedures and facilitate widespread implementation. Following two pilot studies analysing spiked samples and material from 698 patients, we conducted a prospective multicentre cross‐sectional study, including 867 patients treated with rivaroxaban, apixaban or edoxaban in clinical practice to comprehensively evaluate a simple, readily available anti‐Xa assay that would accurately measure drug concentrations and correctly predict relevant levels in clinical practice. Anti‐Xa activity was measured by an assay calibrated with low‐molecular‐weight heparin (LMWH) in addition to ultra‐high performance liquid chromatography‐tandem mass spectrometry (LC‐MS/MS). As an external validation, LMWH‐calibrated anti‐Xa activity was also determined in nine external laboratories. The LMWH‐calibrated anti‐Xa activity correlated strongly with rivaroxaban, apixaban or edoxaban drug levels [ r s = 0·98, 95% confidence interval (CI) 0·98–0·98]. The sensitivity for the clinically relevant cut‐off levels of 30, 50 and 100 µg/l was 96·2% (95% CI 94·4–97·4), 96·4% (95% CI 94·4–97·7) and 96·7% (95% CI 94·3–98·1) respectively. Concordant results were obtained in the external validation study. In conclusion, a universal, LMWH‐calibrated anti‐Xa assay accurately measured rivaroxaban, apixaban and edoxaban concentrations and correctly predicted relevant drug concentrations in clinical practice.
No takes yet. Share an insight, caveat, or question.
Willekens et al. (2021) conducted a cross-sectional in Patients treated with DOACs (rivaroxaban, apixaban, or edoxaban) (n=867). LMWH-calibrated anti-Xa assay vs. Ultra-high performance liquid chromatography-tandem mass spectrometry (LC-MS/MS) was evaluated on Correlation with rivaroxaban, apixaban, or edoxaban drug levels measured by LC-MS/MS (rs = 0.98, 95% CI 0.98-0.98). A universal LMWH-calibrated anti-Xa assay strongly correlated with DOAC drug levels measured by LC-MS/MS (rs = 0.98) and had ≥96.2% sensitivity for clinically relevant cut-offs.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: