Key result
Pulmonary vein ostium diameters and elliptical area correlated negatively with nadir balloon temperature during 28-mm cryoballoon ablation (r=-0.41, P<0.0001).
Why the study?
The ideal pulmonary vein and left atrial diameters for achieving optimal cooling temperatures during cryoballoon ablation were unknown.
What are the optimal pulmonary vein ostium diameters for achieving optimal nadir balloon temperature during 28-mm cryoballoon ablation in patients with atrial fibrillation?
Observational (n=71)
What are the optimal pulmonary vein ostium diameters for achieving optimal nadir balloon temperature during 28-mm cryoballoon ablation in patients with atrial fibrillation?
Specific pulmonary vein ostium dimensions (19 < PVmax ≤ 21 mm, 14 < PVmin ≤ 16 mm) predict optimal nadir balloon temperatures during 28-mm cryoballoon ablation for atrial fibrillation.
PV ostium size may influence cryoballoon nadir temperatures; leaves open whether size-based selection improves AF ablation outcomes.
Background Large pulmonary veins (PVs) have the risk of excessive cooling when the cryoballoon is moved into them, whereas smaller PVs may not be isolated because of insufficient balloon‐tissue contact. We previously reported that the effective nadir balloon temperature (NBT) was <− 44°C for predicting early PV reconnections. However, the ideal PV and left atrial (LA) diameters for cooling temperatures during cryoballoon ablation (CBA) are unknown. Methods We measured the PV ostium (PVos) and LA diameters on computed tomography images in 71 atrial fibrillation patients who underwent a 28‐mm CBA. We defined −60°C ≤ NBT <− 44°C as the optimal NBT. The optimal PVos diameters and elliptical area were determined using the receiver operating characteristic curve cutoff values for the optimal NBT. Results The maximum PVos diameter (PVmax), minimum PVos diameter (PVmin), and PV elliptical area correlated negatively with the NBT (PVmax: r = −.34, P < .0001; PVmin: r = −.41, P < .0001; PV elliptical area: r = −.41, P < .0001). Based on the NBT‐derived definition, the optimal PVmax, PVmin, and PV elliptical area were 19 < PVmax ≤ 21 mm, 14 < PVmin ≤ 16 mm, and 186.83 < PV elliptical area ≤ 254.34 mm 2 , respectively. In addition, a positive correlation between the LA longitudinal diameter and NBT was observed in the right superior PV ( r = .34, P = .004), and the LA transverse diameter correlated inversely with the NBT in the left inferior PV ( r = −.34, P = .004). Conclusions We identified the optimal PV diameters and elliptical area to establish the optimal NBT during the second‐generation 28‐mm CBA. These data will help for the preclinical judgment for a successful CBA.
No takes yet. Share an insight, caveat, or question.
Kurokawa et al. (2019) conducted an observational in Atrial fibrillation (n=71). Pulmonary vein ostium and left atrial diameters was evaluated on Correlation with nadir balloon temperature (NBT) (r = -0.41, p=<0.0001). Pulmonary vein ostium diameters and elliptical area correlated negatively with nadir balloon temperature during 28-mm cryoballoon ablation (r=-0.41, P<0.0001).
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: