Key result
A novel physiologic method for deriving central blood pressure from a standard automatic arm cuff yielded precision errors of 6.8 to 9.0 mmHg, significantly lower than the 8.2 to 12.5 mmHg errors of the conventional oscillometric method.
Why the study?
Does a physiologic method for central BP monitoring via a standard automatic arm cuff improve measurement accuracy compared to conventional oscillometric methods in patients undergoing cardiac catheterization?
Observational (n=87)
No
Does a physiologic method for central BP monitoring via a standard automatic arm cuff improve measurement accuracy compared to conventional oscillometric methods in patients undergoing cardiac catheterization?
p-value: p=≤0.01
May support cuff-based central BP use in select settings; leaves open larger validation before practice change.
Current oscillometric devices for monitoring central blood pressure (BP) maintain the cuff pressure at a constant level to acquire a pulse volume plethysmography (PVP) waveform and calibrate it to brachial BP levels estimated with population average methods. A physiologic method was developed to further advance central BP measurement. A patient-specific method was applied to estimate brachial BP levels from a cuff pressure waveform obtained during conventional deflation via a nonlinear arterial compliance model. A physiologically-inspired method was then employed to extract the PVP waveform from the same waveform via ensemble averaging and calibrate it to the brachial BP levels. A method based on a wave reflection model was thereafter employed to define a variable transfer function, which was applied to the calibrated waveform to derive central BP. This method was evaluated against invasive central BP measurements from patients. The method yielded central systolic, diastolic, and pulse pressure bias and precision errors of -0.6 to 2.6 and 6.8 to 9.0 mmHg. The conventional oscillometric method produced similar bias errors but precision errors of 8.2 to 12.5 mmHg (p ≤ 0.01). The new method can derive central BP more reliably than some current non-invasive devices and in the same way as traditional cuff BP.
No takes yet. Share an insight, caveat, or question.
Natarajan et al. (2017) conducted an observational in Patients admitted for diagnostic cardiac catheterization (n=87). Physiologic method for central blood pressure monitoring vs. Conventional oscillometric method was evaluated on Central systolic, diastolic, and pulse pressure precision errors (p=≤0.01). A novel physiologic method for deriving central blood pressure from a standard automatic arm cuff yielded precision errors of 6.8 to 9.0 mmHg, significantly lower than the 8.2 to 12.5 mmHg errors of the conventional oscillometric method.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: