Key result
In patients with a Medtronic Reveal LINQ implantable cardiac monitor, the mean time between an arrhythmic event and physician notification was 15.3 days, with 63% of delays caused by proximity issues.
Why the study?
Clinical success of ICMs depends on timely data transmission following arrhythmic events, but no reports had evaluated how efficiently ICM data is transmitted.
Observational (n=520)
No
Data transmission from implantable cardiac monitors is frequently delayed by over two weeks due to connectivity and proximity issues, highlighting a critical gap in remote arrhythmia surveillance.
Background: The Implantable Cardiac Monitor (ICM) is an invaluable tool for detecting cardiac arrhythmias by providing physicians. Critical to the success of ICMs depends on how quickly and accurately the data can be transmitted to a physician’s office after an arrhythmic event. Then, the clinical event can be analyzed and the treatment will be provided accordingly. However, no reports have been published as to how efficiently the ICM data is transmitted. Methods: There is a retrospective review of 520 patients who received a Medtronic Reveal LINQTM between 2/01/2015 and 6/01/2017. The time from the arrhythmic event to the time of physician notification was calculated and reason for delay was noted. Results: One hundred and twenty patients out of 520 patients (23%) had arrhythmic events transmitted over a mean follow up of 14 ± 4 months. The mean time between cardiac events and physician notification was 15 ± 8 days. Sixty-three percent (63%) of data transmission delay (defined as >24 hours) was due to the MyCareLinkTM Monitor not being in proximity to the patient. Connection failure between the monitor and the network accounted for 34% of data transmission delay. Conclusion: Significant delay in data transmission from Medtronic Reveal LINQTM cardiac monitor occurs frequently impacting patient care. Newer generations of the implantable cardiac monitors utilize Bluetooth technology, enabling immediate transfer of data from ICM to a patient’s cellular phone and subsequently to their physician’s office. This technology could potentially improve efficiency and reliability eliminating the issues of proximity and connectivity.
No takes yet. Share an insight, caveat, or question.
Chokesuwattanaskul et al. (2019) conducted an observational in Arrhythmia monitoring (n=520). Medtronic Reveal LINQ Implantable Cardiac Monitor was evaluated on Time from arrhythmic event to physician notification. In patients with a Medtronic Reveal LINQ implantable cardiac monitor, the mean time between an arrhythmic event and physician notification was 15.3 days, with 63% of delays caused by proximity issues.