Key result
Legacy Medtronic CIED transceivers deliver ~91% of remote alerts to CareLink within 24 hours.
Why the study?
Alert-based remote monitoring systems notify physicians of issues in near real-time, but their effectiveness is contingent upon device connectivity.
Observational (n=109,179)
Yes
Legacy remote monitoring transceivers provide satisfactory alert transmission times, with over 90% of alerts received within 24 hours, though patient connectivity gaps remain a barrier to real-time functionality.
Background: Remote monitoring of cardiac implantable electric devices improves patient outcomes and experiences. Alert-based systems notify physicians of clinical or device issues in near real-time, but their effectiveness is contingent upon device connectivity. Objective: To assess patient connectivity by analyzing alert transmission times from patient transceivers to the CareLink network. Methods: Alert transmissions were retrospectively gathered from a query of the United States de-identified Medtronic CareLink database. Alert transmission time was defined as the duration from alert occurrence to arrival at the CareLink network and was analyzed by device type, alert event, and alert type. Using data from previous studies, we computed the benefit of daily connectivity checks. Results: The mean alert transmission time was 14.8 hours (median = 6 hours), with 90.9% of alert transmissions received within 24 hours. Implantable pulse generators (17.0 ± 40.2 hours) and cardiac resynchronization therapy-pacemakers (17.2 ± 42.5 hours) had longer alert transmission times than implantable cardioverter-defibrillators (13.7 ± 29.5 hours) and cardiac resynchronization therapy-defibrillators (13.5 ± 30.2 hours), but the median time was 6 hours for all 4 device types. There were differences in alert times between specific alert events. Based on our data and previous studies, daily connectivity checks could improve daily alert transmission success by 8.5% but would require up to nearly 800 additional hours of staff time on any given day. Conclusion: Alert transmission performance from Medtronic devices was satisfactory, with some delays likely underscored by patient connectivity issues. Daily connectivity checks could provide some improvement in transmission success at the expense of increased clinic burden.
No takes yet. Share an insight, caveat, or question.
Cronin et al. (2023) conducted an observational in Patients with Medtronic cardiac implantable electronic devices (CIEDs) (n=109,179). Medtronic legacy 24950/24952 MyCareLink patient transceivers was evaluated on Alert transmission time (duration from alert occurrence to arrival at the CareLink network) (Mean 14.8 hours, median 6.0 hours). Alert transmissions from Medtronic CIEDs using legacy transceivers took a mean of 14.8 hours (median 6.0 hours) to reach the CareLink network, with 90.9% successfully received within 24 hours.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: