Key result
Remote monitoring of cardiac implantable electronic devices was safe and feasible, with 73.7% of scheduled transmissions occurring on time and only 1.8% requiring clinical intervention.
Observational (n=57)
No
Remote monitoring of CIEDs is safe, feasible, and highly preferred by patients in a Southeast Asian setting, demonstrating good compliance and prompt clinical review times.
Answers are grounded in this paper and cite their sources.
Supports remote CIED monitoring feasibility; leaves open need for RCTs to assess clinical outcomes versus in-person care.
INTRODUCTION: Remote monitoring of cardiac implantable electronic devices (CIED) has been shown to improve patient safety and reduce in-office visits. We report our experience with remote monitoring via the Medtronic CareLink(®) network. METHODS: Patients were followed up for six months with scheduled monthly remote monitoring transmissions in addition to routine in-office checks. The efficacy of remote monitoring was evaluated by recording compliance to transmissions, number of device alerts requiring intervention and time from transmission to review. Questionnaires were administered to evaluate the experiences of patients, physicians and medical technicians. RESULTS: A total of 57 patients were enrolled; 16 (28.1%) had permanent pacemakers, 34 (59.6%) had implantable cardioverter defibrillators and 7 (12.3%) had cardiac resynchronisation therapy defibrillators. Overall, of 334 remote transmissions scheduled, 73.7% were on time, 14.5% were overdue and 11.8% were missed. 84.6% of wireless transmissions were on time, compared to 53.8% of non-wireless transmissions. Among all transmissions, 4.4% contained alerts for which physicians were informed and only 1.8% required intervention. 98.6% of remote transmissions were reviewed by the second working day. 73.2% of patients preferred remote monitoring. Physicians agreed that remote transmissions provided information equivalent to in-office checks 97.1% of the time. 77.8% of medical technicians felt that remote monitoring would help the hospital improve patient management. No adverse events were reported. CONCLUSION: Remote monitoring of CIED is safe and feasible. It has possible benefits to patient safety through earlier detection of arrhythmias or device malfunction, permitting earlier intervention. Wireless remote monitoring, in particular, may improve compliance to device monitoring. Patients may prefer remote monitoring due to possible improvements in quality of life.
No takes yet. Share an insight, caveat, or question.
Lim et al. (2016) conducted an observational in Patients with cardiac implantable electronic devices (CIED) (n=57). Remote monitoring via Medtronic CareLink network was evaluated on Compliance to scheduled remote transmissions (percentage on time). Remote monitoring of cardiac implantable electronic devices was safe and feasible, with 73.7% of scheduled transmissions occurring on time and only 1.8% requiring clinical intervention.
Synapse has enriched 10 related papers. Consider them for comparative context: