Key result
A two-stage method based on time frequency blind source separation and wavelet de-noising successfully extracted clear fetal electrocardiogram signals from real-world data.
Answers are grounded in this paper and cite their sources.
A novel two-stage signal processing method combining time frequency blind source separation and wavelet de-noising effectively extracts fetal ECG from maternal signals and noise.
Extraction of the fetal electrocardiogram (FECG) from the mixture of maternal electrocardiogram (MECG) and noise is an important but hard task. The electrocardiograph (ECG) signal presents non-stationary. In this work, a two stage method is proposed based on time frequency blind source separation and wavelet de-noising to extract FECG. In the first stage the recorded signals from the electrodes of pregnant woman are prewhitened, then the spatial time frequency matrix are computed using Cohen's class spatial time-frequency distributions, joint diagonalization of the sets of spatial time frequency matrixes, the mixing matrix is estimated and the FECG corrupted by noise is extracted. In the second stage, the corrupted FECG is performed by wavelet transform to eliminate the affect of noise. By the two stages, a clear FECG signal is obtained and experiments on real world data show the validation of the proposed method.
No takes yet. Share an insight, caveat, or question.
Ma et al. (2009) studied Fetal electrocardiogram extraction. Two-stage method based on time frequency blind source separation and wavelet de-noising was evaluated on Extraction of clear FECG signal. A two-stage method based on time frequency blind source separation and wavelet de-noising successfully extracted clear fetal electrocardiogram signals from real-world data.
Synapse has enriched 8 closely related papers on similar clinical questions. Consider them for comparative context: