Key result
A battery-free passive RFID tag system successfully measured heart rates from 30 to 300 BPM up to ten feet away with a standard deviation of less than one beat per minute.
A novel battery-free monitoring system using a passive RFID tag can accurately transmit heart rate data derived from an ECG signal over short distances.
Answers are grounded in this paper and cite their sources.
May enable battery-free remote HR monitoring; leaves open clinical validation and safety testing.
In this paper, we propose a monitoring system that employs a passive RFID tag to transmit heart rate using an ECG signal as its source. This system operates without a battery and has been constructed with easily available commercial components. Here, an RFID tag is used as an on-off keying device, wherein it is normally transmitting, but turns off every time a heart beat is detected. Heart beats ranging from 30BPM through 300BPM are successfully measured using our device. It is shown that the system is capable of providing accurate heart rate measurements up to a distance of ten feet with a standard deviation of less than one beat per minute without a local power source. The proposed system is also found to be resilient in the presence of an additional RFID tag.
No takes yet. Share an insight, caveat, or question.
Vora et al. (2015) studied Heart rate monitoring. Passive RFID tag based heart rate monitoring system was evaluated on Heart rate measurement accuracy and range. A battery-free passive RFID tag system successfully measured heart rates from 30 to 300 BPM up to ten feet away with a standard deviation of less than one beat per minute.
Synapse has enriched 8 closely related papers on similar clinical questions. Consider them for comparative context: