Why the study?
Optimal timing of contrast-enhanced CMR pulmonary vein angiography is important for high image quality before catheter ablation, requiring identification of cardiovascular parameters predicting contrast agent travel time.
Can routine cardiovascular parameters predict optimal contrast agent travel time for 3D CMR pulmonary vein angiography in patients undergoing catheter ablation for atrial fibrillation?
Can routine cardiovascular parameters predict optimal contrast agent travel time for 3D CMR pulmonary vein angiography in patients undergoing catheter ablation for atrial fibrillation?
Routine cardiovascular parameters poorly predict contrast agent travel time, demonstrating that 3D pulmonary vein CMR angiography prior to ablation requires real-time bolus tracking to ensure consistently high image quality.
Routine parameters should not guide contrast timing for pre-ablation CMR; leaves open whether real-time tracking or refined models ensure optimal quality.
BACKGROUND: To achieve high image quality of cardiovascular magnetic resonance (CMR) pulmonary vein (PV) angiography prior catheter ablation in patients with atrial fibrillation, optimal timing of the angiographic sequence during contrast agent passage is important. The present study identified influential cardiovascular parameters for prediction of contrast agent travel time. METHODS: One hundred six consecutive patients underwent a CMR examination including three-dimensional (3D) contrast-enhanced PV angiography with real-time bolus tracking prior to catheter ablation. Correct scan timing was characterized by relative signal enhancement measurements in the pulmonary artery, left atrium (LA), and ascending aorta. Furthermore, left- and right-ventricular function, left- and right-atrial dimensions, presence of mitral or tricuspid insufficiencies, and main pulmonary artery diameter were determined. RESULTS: The highest relative signal enhancement in LA demonstrated optimal scan timing. Contrast agent travel time showed wide variability (range: 12-42 s; mean: 18 ± 4 s). On univariate analysis, most cardiovascular parameters correlated with contrast agent travel time while on multivariate analysis left- and right-ventricular function remained the only independent predictors, but overall a poor fit to the data (adjusted R2, 27.5%) was found. CONCLUSIONS: Contrast agent travel time was mainly influenced by left- and right-ventricular function but prediction models poorly fitted the data. Thus, 3D PV angiography prior to PV ablation procedures necessitates real-time assessment, with visual determination of individual contrast agent passage time to ensure consistently high CMR image quality.
No takes yet. Share an insight, caveat, or question.
Löbe et al. (2020) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: