Why the study?
Does a modified-tip NaviStar catheter improve the amplitude of monophasic action potential recordings compared to an ordinary catheter in swine and humans?
Does a modified-tip NaviStar catheter improve the amplitude of monophasic action potential recordings compared to an ordinary catheter in swine and humans?
A modified-tip NaviStar catheter provides significantly higher monophasic action potential amplitudes compared to an ordinary catheter, demonstrating safety and feasibility for clinical electroanatomic mapping.
Answers are grounded in this paper and cite their sources.
Supports feasibility of modified-tip catheter for MAP recordings; leaves open clinical benefits in electroanatomic mapping.
OBJECTIVE: To evaluate the feasibility of monophasic action potential (MAP) mapping using a modified-tip NaviStar catheter in swine and humans. METHODS: MAP mapping was performed using the modified-tip catheter at 71 +/- 21 atrial and 60 +/- 16 ventricular sites in 10 healthy pigs and at 56 ventricular sites in one patient, and using an ordinary Navi-Star catheter at 30 atrial sites in one patient and 50 +/- 14 ventricular sites in four patients. In an additional 20 patients, MAPs were also recorded at 9 +/- 2 atrial sites using the modified-tip catheter or at 12 +/- 9 atrial sites using the ordinary catheter. RESULTS: In pigs, the plateau amplitudes of the MAPs recorded using the modified-tip catheter were 4.1 +/- 3.2 mV for the atrial and 9.5 +/- 4.3 mV for the ventricular MAPs. In patients, both the ventricular and atrial MAPs recorded using the modified-tip catheter were significantly higher than using the ordinary catheters, 15.7 +/- 8 and 3.0 +/- 0.9 mV vs 9.5 +/- 3.9 and 2.0 +/- 0.6 mV for the ventricular and atrial MAPs, respectively (p < 0.0001). The baseline disturbances were <10% of the MAP amplitude in 95% of the pig and 96% of the patient MAPs. CONCLUSION: A modified-tip Navi-Star catheter could be used in swine and in humans for prompt recording of MAPs with acceptable amplitudes and baselines. MAP mapping using the modified-tip catheter is safe and feasible for clinical use.
No takes yet. Share an insight, caveat, or question.
Liu et al. (2002) studied this question.