Why the study?
Does 1.5 Tesla MRI scanning in patients with non-MRI conditional cardiac devices have similar safety and lead characteristic changes compared to those with MRI conditional devices?
Cohort (n=105)
No
Does 1.5 Tesla MRI scanning in patients with non-MRI conditional cardiac devices have similar safety and lead characteristic changes compared to those with MRI conditional devices?
1.5 T MRI scanning can be performed with similarly low clinical risk in patients with non-MRI conditional cardiac devices compared to those with MRI conditional devices.
May support 1.5T MRI in non-conditional devices; hypothesis-generating and should not yet change practice.
OBJECTIVES: This study compared risks associated with magnetic resonance imaging (MRI) in patients with non-MRI conditional and MRI conditional pacing and defibrillator systems with particular attention to clinically actionable outcomes. BACKGROUND: While recipients of new MRI conditional pacemaker and defibrillator systems may undergo MRI scanning with very low risk, safety and regulatory concerns persist regarding such scanning in recipients of non-MRI conditional systems. METHODS: Patients with any cardiac device who were referred for MRI were prospectively enrolled at a single center and underwent scanning at 1.5 Tesla. Pre- and postscan lead characteristic changes, system integrity, and symptoms were analyzed. A comparison was made between non-MRI conditional and MRI conditional devices. RESULTS: 105 patients were evaluated allowing for comparison of 97 scans with non-MRI conditional devices and 16 scans with MRI conditional devices. The cohort included those with pacemaker dependency, defibrillator, and cardiac resynchronization devices. Small, nonsignificant changes were observed in lead characteristics following scanning, and there was no significant difference when comparing non-MRI and MRI conditional devices. Lead parameter changes did not require lead revision or programming changes. No device reset, failures, or premature scan termination was observed. CONCLUSIONS: 1.5 T MRI scanning in patients with MRI conditional and non-MRI conditional cardiac devices was performed with similar, low clinical risk.
No takes yet. Share an insight, caveat, or question.
Shah et al. (2017) conducted a cohort in Cardiac implantable electronic devices (n=105). Non-MRI conditional cardiac devices vs. MRI conditional cardiac devices was evaluated on Pre- and postscan lead characteristic changes, system integrity, and symptoms.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: