Electrospray ionization with an ultralow flow rate (</=4 nanoliters per minute) was used to directly couple capillary electrophoresis with tandem mass spectrometry for the analysis and identification of biomolecules in mixtures. A Fourier transform mass spectrometer provided full spectra (>30 kilodaltons) at a resolving power of approximately 60,000 for injections of 0.7 x 10(-18) to 3 x 10(-18) mole of 8- to 29-kilodalton proteins with errors of <1 dalton in molecular mass. Using a crude isolate from human blood, a value of 28,780.6 daltons (calculated, 28,780.4 daltons) was measured for carbonic anhydrase, representing 1 percent by weight of the protein in a single red blood cell. Dissociation of molecular ions from 9 x 10(-18) mole of carbonic anhydrase gave nine sequence-specific fragment ions, more data than required for unique retrieval of this enzyme from the protein database.
No takes yet. Share an insight, caveat, or question.
Valaskovic et al. (1996) studied this question.